17. Image, Multiple, and Out of Rand Responses Test
Image and out-of-band responses are checked by setting the analyzer
center frequency to several frequencies across the analyzer range and
tuning a leveled signal source to the frequencies determined by the
tuning equation, Fsiz = nFLof F IF. Input signals at these frequencies
will excite all possible image and out-of-band responses for a given
1st LO frequency and all positive integer values of n. In this test,
only values of n corresponding to the analyzer mixing modes are
used. Multiple responses are checked by applying an input signal
and measuring the response at those center frequencies for which a
harmonic of the 1st LO mixes with the input signal.
S P E C T R U M A N A L Y Z E R
ADAPTER
Figure 2-31. Image, Multiple, and Out-of-Rand Responses Test Setup
Synthesized Sweeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8340A
61 cm (24 in.) Cable Assembly, SMA Male Connectors .... 5061-1086
Adapter, Type N (m) to SMA (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1250
Adapter, SMA Female Connectors . . . . . . . . . . . . . . . . . . . . . . . . . .1250-1158
BNC Tee . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-0781
BNC
Performance Tests 2-73