Frequency Response Adjustment - HP Q METER 4342A Operating And Service Manual

Table of Contents

Advertisement

Model 4342A
d.
Connect
the
Test
Oscillator
to the
4342A as shown in Figure
5-6.
e.
Set the
Test
Oscillator
frequency
to
1OOkHz and output
for
full
scale
reading
(approx.
9OOmVrms) on 4342A Q
meter.
f.
Adjust
A6R6 (REC GAIN adj.)
for
1V
?O.OlV on Digital
Voltmeter
display.
g.
Repeat
steps
c through
f because
both
adjustments
interact.
5-23.
FREQUENCY RESPONSE ADJUSTMENT.
An RF Oscillator
and an RF Voltmeter
(with
known frequency
flatness)
are required
for
this
adjustment.
a.
Connect
RF Oscillator
and RF Voltmeter
as shown in Figure
5-7.
b.
Set 4342A C and AC dials
to minimum.
C.
Set RF Oscillator
frequency
to
1OMHz
and its
output
for
full
scale
meter
deflection
(approx.
9OOmVrms) on 4342A
Q meter.
d.
Note
RF Voltmeter
reading.
e.
Set RF Oscillator
frequency
to 20MHz
and its
output
for
the
same RF Volt-
meter
reading
as that
noted
in step
d.
Section
V
Paragraphs
5-23 and 5-24
f.
Adjust
A5R32 (MED. FREQ. ADJ) for
full
scale
reading
on 4342A Q meter.
g.
Set RF Oscillator
frequency
to
7OMHz
and its
output
for
the
same RF Volt-
meter
reading
as that
noted
in
step
d.
h.
Adjust
A5C16 (HIGH FREQ. ADJ) for
full
scale
reading
on 4342A Q meter.
1.
Repeat
steps
c through
h until
both
difference
(from
full
scale)
Q meter
readings
obtained
in steps
f and h are
within
+2% of
full
scale.
5-24.
Q LIMIT
SELECTOR ADJUSTMENT.
An RF Oscillator
is required
for
this
adjust-
ment.
a.
b.
C.
d.
e.
Set 4342A rear
panel
OVER LIMIT
DIS-
PLAY TIME switch
toooposition.
Connect
RF Oscillator
between
HI and
GND terminals.
Set 4342A Q LIMIT
control
to 100.
Set RF Oscillator
to desired
frequency
(100kHz
to 1MHz) and adjust
its
output
for
full
scale
reading
on 4342A Q
meter.
Rotate
A7R3 (Q-PRESET adj.)
CCW until
front
panel
OVER LIMIT
indicator
lights.
RF OSCILLATOR
Figure
5-7.
Frequency
Response
Adjustment.
5-11