Down/Up Converter Adjustments; Down/Up Converter Adjustments Setup - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

IF-Display Section
Resolution Bandwidth Switching Uncertainty Test
The CAL OUTPUT signal is connected to the RF INPUT connector of
the spectrum analyzer and controls are set to display the signal in a
narrow bandwidth. A marker is placed at the peak of
measure the peak amplitude. The bandwidth is changed to a wide
bandwidth and the Down/Up Converter is adjusted to place the peak
amplitude of the signal
signal. Optionally, the input signal is removed and the IF signal is
the
monitored at
output of the Bandwidth Filters using a spectrum
analyzer with an active probe. The 18.4 MHz Local Oscillator and all
harmonics are then adjusted for minimum amplitude.
Figure 3-50. Down/Up Converter Adjustments Setup
Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8566B
Active Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 85024A
Probe Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 1122A
Adjustment A4A6AlR29 WIDE GAIN adjusts the amplitude of the
21.4 MHz Bandwidth Filters (3 kHz through 3 MHz) relative to the
amplitude of the 3 MHz Bandwidth Filters (1 kHz through 10 Hz).
Perform Adjustment Procedure 6, "Log Amplifier Adjustments,"
(steps 1-12) Adjustment Procedure 8, "21.4 MHz Bandwidth Filter
Adjustments," and Adjustment Procedure 7, "3 MHz Bandwidth Filter
Adjustments" to ensure that the amplitudes of the bandwidth filters
are optimized before adjusting A4A6AlR29 WIDE GAIN (steps 14-17).
the same
as the level of the narrow bandwidth
the
signal to