Resolution Bandwidth Accuracy Test; Resolution Bandwidth Measurement; Bandwidth Accuracy - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

Figure 2-5. Resolution Bandwidth Measurement
8. Vary spectrum analyzer settings according to 'Ihble 2-5. Press
SWEEP (SINGLE) and measure the 3 dB bandwidth for each
resolution bandwidth setting by the procedure of steps 6 and 7 and
record the value in Table 2-5. The measured bandwidth should fall
between the limits shown in the table.
SPAN
3 MHz
5 MHz
1 MHz
2 MHz
300 kHz
500 kHz
100 kHz
200 kHz
30 kHz
50 kHz
10 kHz
20 kHz
3 kHz
5 kHz
1 kHz
2 kHz
300 Hz
500 Hz
100 Hz
200 Hz
30 Hz
100 Hz
10 Hz
100 Hz

3. Resolution Bandwidth Accuracy Test

MARKER A Readout of 3 dB Bandwidth 1
]
Actual
Min
900 kHz
270.0 kHz
90.0 kHz
27.00 kHz
9.00 kHz
2.700 kHz
800 Hz
240 Hz
80 Hz
24.0 Hz
8.0 Hz
3.600 MHz
1.100 MHz
330.0 kHz
110.0 kHz
33.00 kHz
11.00 kHz
3.300 kHz
1.200 kHz
360 Hz
120 Hz
36.0 Hz
12.0 Hz
Performance Tests 2-13