Residual Responses Test - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

The spectrum analyzer is tested for residual responses across its
frequency range with no signal applied and 0 dB input attenuation.
50 Ohm Coaxial Termination . . . . . . . . . . . . . . . . . . HP 909A, Option 012
1.
Press 1-j.
2.
Connect CAL OUTPUT to RF INPUT. Press (RECALL) @J.
3.
Adjust AMPTD CAL for a MARKER amplitude of -10.00 dBm
4. Press (O]. Disconnect CAL OUTPUT and terminate RF
INPUT with a 50 ohm coaxial termination.
5. Key in the following:
]
. . . . . . . . . . . . . . . . . . . -20 dBm
REFERENCE LEVEL
. . . . . . . . . . . . . . . . . . . . . . . . 30 kHz
(VIDEO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 kHz
Press DISPLAY LINE [ENTER] and key in -100 dBm.
6.
There should be at least 3 dB margin between the noise trace and
the display line so that any residual responses may be distinguished
from the noise. It may be necessary to reduce the resolution or video
bandwidths from the settings given in this procedure to achieve this
margin. If this causes the MEAS UNCAL message to appear, it will be
necessary to reduce the frequency span and use more sweeps to cover
the frequency range.
(
7. Press TRACE A
CLEAR
completion of the sweep. (See Figure 2-27.)

15. Residual Responses Test

-
),
SWEEP [SINGLEj. Wait for
WRITE
Performance Tests 2-61