10. Frequency Response Test
2-36 Performance Tests
)
18. Set
to 100 Hz. Step analyzer
STEP SIZE
from 100 Hz to 1 kHz with 0J and set function generator
FREQUENCY to center signal on display at each step. Press
MARKER
SEARCH
should be -2 dBm f0.6 dB.
Deviation 100 Hz to 1 kHz
19. Connect equipment as shown in Figure 2-16.
SPECTRUM ANALYZER

Frequency Response Test Setup (100 kHz to 60 MHz)

20. Press @7ZX) on the analyzer. Key in the following settings:
[
T
CEN
ER
FREQUENCY
] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
(
FREQUENCY SPAN
C
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 dB
ENTER
21. Set the controls of the frequency synthesizer as follows:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 MHz
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -2 dBm
22. Adjust the output amplitude of the frequency synthesizer to place
the signal at the 8th graticule line.
23. Key in the analyzer settings as follows:
(
START
24. Key in the frequency synthesizer settings as follows:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . .2,000,100 Hz
SWEEP WIDTH . . . . . . . . . . . . . . . . . . . . . . . 3,998,OOO Hz
[
25. Press TRACE A
CLEAR
(
)
at each frequency. The MKR amplitude
FREQUENCY SYNTHESIZER
Figure 2-16.
)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
-
]
and [MAX) on the analyzer.
WRITE
)
CENTER FREQUENCY
MHZ
MHz
.O
Hz