20. Sweep + Tune Out Accuracy Test

CENTER
Voltmeter Reading
FREQUENCY
Min
0 Hz
-0.010
1 MHz
-0.011
1 2
MHz
-0.022
130 MHz
-0.143
670 MHz
-0.693
1.3 GHz
5.7 GHz
-5.824
12.5 GHz -12.760
18.6 GHz -18.982
22 GHz -22.450
(Volts)
Actual
I
+ 0.010
+ 0.009
-0.002
-0.117
-0.647
-1.264
-5.576
-12.240
-18.218
-21.550