Line-Related Sidebands Test; Line Related Sidebands Test Setup - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

I
Offset from Carrier Center Frequency Sidebands
360 Hz to 2 kHz
Option 400:
For center frequencies >lOO Hz and ~5.8 GHz:
Offset from Carrier Center Frequency Sidebands
The line-related sidebands are measured on signals of 100 MHz, 2.4
is used as the signal source for this test. Therefore, if measured
values are not within specified limits, either analyzer may be at fault.
S P E C T R U N A N A L Y Z E R
Figure 2-24. Line Related Sidebands Test Setup

13. Line-Related Sidebands Test

I
100 MHz to 5.8 GHz
-60 dBc I
100 Hz to 100 MHz I - 7 0 dBc
A D A P T E R
Performance Tests 2-53
I
A N A L Y Z E R
A D A P T E R
A D A P T E R
I i