Adjustable Components - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

Reference
Adjustment
Designator
Name
ZHF GAIN
INT GAIN
FOCUS COMP
ORTHO
3 D
INTENSITY
LIMIT
ASTIG
FOCUS LIMIT
X POSN
X GAIN
XHF GAIN
Cl04
Cl09
Y POSN
Y GAIN
YHF GAIN
+ 15 ADJ
HV ADJUST
Serial Prefix 3001A and below,
exceptions to AlA through AlA6.
Adjustment
Number
3
Adjusts rise and fall times of Z axis amplifier pulse.
3
Adjusts rise and fall times of Z axis amplifier pulse.
Sets adjustment range of front-panel INTENSITY
3
control.
3
Corrects focus for beam intensity.
Magnitude of top/bottom focus correction.
Centering of top/bottom focus correction.
Magnitude of right/left focus correction.
Centering of right/left focus correction.
3
Sets orthogonality of CRT.
3
Adjusts spot size.
3
Sets adjustment range of front-panel INTENSITY
control.
3
Adjusts astigmatism of CRT.
3
Coarse adjusts CRT focus.
3
Adjusts rise and fall times of X deflection amplifier
pulse.
3
Adjusts rise and fall times of X deflection amplifier
pulse.
Adjusts horizontal position of trace.
3
Adjusts horizontal gain of trace.
3
Adjusts rise and fall times or X deflection amplifier
pulse.
Adjusts rise and fall times of Y deflection amplifier
3
pulse.
Adjusts rise and fall times of Y deflection amplifier
3
pulse.
Adjusts vertical position of trace.
Adjusts vertical gain of trace.
Adjusts rise and fall times of Y deflection amplifier
pulse.
1
Adjusts + 15 V dc supply voltage.
2
Adjusts CRT high voltage.
see back of table for
Adjustment Function
Adjustments 3-5