Sample And Hold Balance Adjustment Waveforms; Waveform Before Adjustment - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

- 2 . 9 0 0 0 0 us
Figure 3-106. Sample and Hold Balance Adjustment Waveforms
-
- 2 . 9 0 0 0 0 us
Figure 3-107. Waveform Before Adjustment
19. Connect the oscilloscope Channel 1 probe to A3A3TP7.
20. Adjust A3A2R51 Y S&H BAL for minimum dc offset between the
level of the signal inside the two pulses to the signal level outside
the two pulses.
21. Set spectrum analyzer LINE switch to STANDBY.
22. Reinstall A3A3 Line Generator in spectrum analyzer without
extender boards.
23. Set spectrum analyzer LINE
24. Digital Storage Display Adjustments
2 . 1 0 0 0 0 us
- 4 0 0 . 0 0 0 ns
5 0 0
- 4 0 0 . 0 0 0 ns
2 . 1 0 0 0 0 us
5 0 0
switch to
ON.
1
5 0 . 0 nV/div
10.00
:
dc
f
3 0 0 . 5
1
5 0 . 0 W/div
10.00
:
dc
300.5 l
l
Adjustments 3-215