Typical Coarse Frequency Response (10 Mhz - 2.5 Ghz) - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

32. On the spectrum analyzer, key in 1-1, (SHIFT] TRACE A
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 MHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 GHz
RESBW.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3MHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -7 dBm
LOG SCALE
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2s
33. Set the synthesized sweeper controls as follows:
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 MHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 GHz
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .INT
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 ms
SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CONT
34. On the spectrum analyzer, key in TRACE A
(SWEEP TIME) 2S, MARKER (j-1 500 MHZ, IHOLD). AS
the spectrum analyzer completes each sweep, a series of
approximately 18 new responses should be displayed, as shown
in Figure 3-92. The peaks of these responses coarsely outline the
spectrum analyzer frequency response.

Typical Coarse Frequency Response (10 MHz - 2.5 GHz)

35. Gradually readjust (Band A Lower Segment) A6AllR48 Al and
(Band A Upper Segment) A6AllR66 A2 for maximum flatness of
the displayed signal responses. The adjustments are interactive,
with A6AllR48 Al having the most effect on the level of the
displayed signals below approximately 500 MHz (the marker
position), and A6AllR66 A2 having the
of the displayed signals above approximately 500 MHz. Adjust
displayed signal responses below 500 MHz. Adjust A6AllR66 A2
2 1. Frequency Response Adjustments
@
LEAR
Figure 3-92.
most effect
Adjustments 3-181
dB/DIV
-
),
WRITE
on the level

Advertisement

Table of Contents
loading

Table of Contents