Narrow Span Test Setup - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

2-8 Performance Tests
Sweep, DAC, and Main Coil Driver Adjustments
For spans In X 5 MHz, fl% of indicated frequency separation.
For spans >n X 5 MHz, f 3% of indicated frequency separation.
1
2 5.8 GHz to 12.5 GHz
3 12.5 GHz to 18.6 GHz
4 >18.6 GHz
* n is the harmonic mixing number, depending on center
frequency.
Spans less than 100 MHz are checked
by comparing the displayed frequency span of two signals with their
known span. Wider spans are tested by tuning a synthesized sweeper
from one edge of the analyzer display to the other and measuring the
frequency change with a frequency counter.
Figure 2-3. Narrow Span Test Setup
Equipment listed is for two test setups, Figure 2-3 and Figure 2-4.
100 Hz to 5.8 GHz
with
a frequency synthesizer
SPECTRUM ANALYZER