13. Sweep, DAC, and Main Coil Driver Adjustments
Al6
A19
SCAN
DAC
GENERATOR
TOP VIEW
Figure 3-58. Location of Sweep and DAC Adjustments
10. Adjust A16R67 SWEEP TIME adjustment for sweep ramp of 500
ms duration (not including dead time at beginning and end of
each ramp) as measured on the oscilloscope.
11. Press
12. Connect the oscilloscope channel 1 to A16TP4 (AUX OUT). Display
should be a stepped series of 4 sweep ramps similar to that shown
in Figure 3-59.