Test 17. Image, Multiple, And Out-Of-Band Responses - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

Step 8. Image and Out-of-Rand Response
Spectrum Analyzer Synthesized Sweeper
(
FREQUENCY)
CENTER
3
3642.800
6321.400
6964.200
6
2517.900
3160.700
5357.200
9
4017.900
4660.700
8357.200
12696.500
13339.300
12
5517.900
6160.700
11357.200
17196.500
17839.300
4571.500
15
5214.300
9464.300
10107.100
14357.200
17
5238.100
5880.900
10797.700
11440.500
16357.200
Displayed Spurious
Frequency
Measured
Maximum
-70
-60
-60
-60
-60
-70
-60
-60
-70
-60
-60
-60
-60
-70
-60
- 6 0
-60
-60
-60
- 6 0
-70
-60
- 6 0
-60
-60
-70
Performance Tests 2-109