Frequency Response Measurement (60 MHz to 2.5 GHz)

48.
On the spectrum analyzer, key in TRACE B LVIEW), TRACE A
a marker on the highest point on the TRACE B waveform.
Record the amplitude and frequency for this maximum point in
-8.80 dBm.
49.
On the spectrum analyzer, use the DATA knob to position
a marker on the lowest point on the TRACE B waveform.
Record the amplitude and frequency for this minimum point in
than - 11.20 dBm.
50.
On the spectrum analyzer, press MARKER [al and use the DATA
knob to position a second marker
TRACE B waveform. Flatness (total peak-to-peak amplitude
deviation) of the displayed trace should be less than 1.20 dB.
On the spectrum analyzer, key in
51.
(KS=), ISHIFT) TRACE A C-1 (KSb). Set the spectrum
analyzer controls as follows:
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.0 GHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.9 GHz
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 MHz
VIDEO BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 MHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -7 dBm
LOG SCALE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 dB/DIV
52.
Consult the power sensor Cal Factor versus Frequency graph or
table and set the power meter CAL FACTOR % switch to the 3
10. Frequency Response Test
,
I
I
V B W 3 M H z
Figure 2-20.
on
the highest point on the
S T O P 2 . 5 0 GHz
S W P 6 2 . 3 rnsec
)
PEAK
Performance Tests 2-41