13. Sweep, DAC, and Main Coil Driver Adjustments
1
2 . 0 0
10.00
:
I
dc
- 1 . 0 0 0 0 0 s
0 . 0 0 0 0 0 s
1 . 0 0 0 0 0 s
2 0 0
1.375
v
Figure 3-59. Properly Adjusted DC Levels Between Sweep Ramps
13. Adjust A16R68 AUX adjustment to align dc level of 3 dead time
steps with upper dc level of each preceding sweep ramp. Refer
to Figure 3-59 and Figure 3-60 for typical display of proper and
improper adjustment.
1
2 . 0 0
10.00
:
I
dc
2 0 0
1.375
v
Figure 3-60. Improperly Adjusted DC Levels Between Sweep Ramps
14. Disconnect the oscilloscope (and universal counter) from the
spectrum analyzer.
Adjustments 3-l 17