Frequency Response - HP 8566B Installation And Verification Manual

Spectrum analyzer, includes option 400, option 462, and option 857
Hide thumbs Also See for 8566B:
Table of Contents

Advertisement

14. Frequency
Response
Specification
Related Performance
Description
5-24 Operation Verification
Center Frequency
Non-Preselected Band
200 Hz to 2.5 GHz
Preselected band
2.0 GHz to 5.8 GHz
5.8 GHz to 12.5 GHz
2.5 GHz to 18.6 GHz
18.6 GHz to 20.0 GHz
20.0 GHz to 22.0 GHz
Cumulative Flatness
200 Hz to 22.0 GHz
Frequency Response Test
If this test is individually selected, a menu of the testable bands is
displayed. For all bands, the test consists of 100 data points taken
across the selected band. For the 200 Hz to 22 GHz test, six bands are
tested (600 data points). First the sweeper is used to test from 20 MHz
to 22 GHz, and then the synthesizer is used to test from 200 Hz to 20
MHz. The levels at 20 MHz are matched to guarantee continuity. If
the Frequency Response test is entered from the ALL TESTS Mode,
then the 200 Hz to 22 GHz is selected automatically.
Flatness
f1.7 dB (3.4 dB)
f1.7 dB (3.4 dB)
f2.2 dB (4.4 dB)
f2.2 dB (4.4 dB)
f3.0 dB (6.0 dB)
6.0 dB