AND AD-4430B Instruction Manual page 28

Din rail weighing module
Hide thumbs Also See for AD-4430B:
Table of Contents

Advertisement

Function No.
Parameter
Function name
and range
CAL08
Stability detection
0.0 to 9.9
CAL09
Stability detection
0 to 9
CAL09
Weight value
STABLE signal
Tare and zero
CAL10
adjustment when
0 to 1
CAL11
Tare when the gross
0 to 1
weight is negative
Output when
CAL12
overflow and
0 to 1
CAL13
Exceeding negative
1 to 3
gross weight
CAL14
Exceeding negative
1 to 2
net weight
CAL15
Clear the zero value
0 to 1
Performs stability detection using this setting in
combination with the setting of the stability detection
band CAL09. When 0.0, stability detection will not
time
be performed. (Stable all the time) Unit : Second in
0.1 increments.
Performs stability detection using this setting in
combination with the setting of the stability detection
time CAL08. When 0, stability detection will not be
band
performed. (Stable all the time) Unit : d (minimum
division).
CAL08
Tare and zero adjustment when the weight value is
unstable.
0 : Disables both functions.
unstable
1 : Enables both functions.
Tare when the gross weight is negative.
0 : Disables tare.
1 : Enables tare.
Standard serial output when the weight value
overflows and is unstable.
0 : Disables output.
unstable
1 : Enables output.
To judge when the negative gross weight is exceeded.
1 : Gross weight < -99999
2 : Gross weight < Negative weighing capacity
3 : Gross weight < -19d
To judge when the negative net weight is exceeded.
1 : Net weight < -99999
2 : Net weight < Negative weighing capacity
Select whether or not to clear the zero value.
0 : Disables.
1 : Enables.
Description
Stability detection outputs the STABLE
signal when changes in the weight
value are within a certain range during
a certain time.
CAL08
- 26 -
Default
value
1.0
2
CAL09
Time
1
1
1
1
1
1
AD-4430B