7.6.4.

The List Of The Calibration

Function No.
Function
Parameter
name
and range
CAL02
Decimal point
0 to 0.0000
position
CAL03
Minimum
1 to 50
division
CAL04
Weighing
1 to 99999
capacity
Zero
CAL05
adjustment
0 to 100
range
CAL06
Zero tracking
0.0 to 5.0
time
CAL07
Zero tracking
0.0 to 9.9
band
CAL08
Stability
0.0 to 9.9
detection time
Stability
CAL09
detection
0 to 9
band
Tare and zero
CAL10
adjustment
0 to 1
when
unstable
Tare when
CAL11
the gross
0 to 1
weight is
negative
Description
Decimal point position of the weight value
0
0.0
0.000
Minimum division (d) of the weight value
1
2
5
10
Weighing capacity of the module.
Weighing is possible up to the value of this
setting plus 8 d. If the value exceeds this,
overflow will occur and will not be displayed. The
decimal point position is the same as the setting
of CAL02.
Range to enable zero adjustment by the
key Expressed as a percentage of
(ZERO)
the weighing capacity with the calibration zero
point as the center. For example, if this is set to 2,
the value in the range of ±2% of the weighing
capacity with the calibration zero point at the
center will be to zero. When a power-ON zero is
performed, the initial zero point will be the center.
Performs zero tracking using this setting in
combination with the setting of the zero tracking
band CAL07. When 0.0, zero tracking will not be
performed. Unit : Second in 0.1 increments.
Performs zero tracking using this setting in
combination with the setting of the zero tracking
time CAL06. When 0.0, zero tracking will not be
performed. Unit : d (minimum division) in 0.1
increments.
Performs stability detection using this setting in
combination with the setting of the stability
detection band CAL09. When 0.0, stability
detection will not be performed. (Stable all the
time) Unit : Second in 0.1 increments.
Performs stability detection using this setting in
combination with the setting of the stability
detection time CAL08. When 0, stability
detection will not be performed. (Stable all the
time) Unit : d (minimum division).
Tare and zero adjustment when the weight
value is unstable.
0 : Disables both functions.
1 : Enables both functions.
Tare when the gross weight is negative.
0 : Disables tare.
1 : Enables tare.
- 46 -
0.00
0.0000
20
50
Default
User
MEMO
value
setting
0
1
70000
2
0.0
0.0
1.0
2
1
1
AD-4430B