AND AD-4430B Instruction Manual page 29

Din rail weighing module
Hide thumbs Also See for AD-4430B:
Table of Contents

Advertisement

Function No.
Parameter
Function name
and range
CAL16
Zero setting when
0 to 1
power is turned on
CAL17
-7.0000
Input voltage at zero
to 7.0000
CAL18
0.0001
Input voltage at span
to 9.9999
Calibration weight
CAL19
value corresponding
1 to 99999
to Input voltage at
NOTE:
*1
Record the setting values of CAL17,
manual to prepare against a failure.
*2
By changing the parameters of CAL17,
calibration" can be adjusted optionally. (Digital span accuracy approximately 1/5000. The
accuracy varies depending on the load cell output accuracy and the conditions of calibration.)
Except for an emergency, perform calibration with an actual load.
CAL26
Gravity acceleration
9.7500
of place of calibration
to 9.8500
CAL27
Gravity acceleration
9.7500
of place of use
to 9.8500
CAL28
Disable averaging
0 to 1
hold
AD-4430B
Select whether or not to perform zero setting when
power is turned on.
0 : Disables.
1 : Enables.
Input voltage from a load cell at zero. Unit : mV/V.
This value is determined in zero calibration during
the calibration with an actual load.
Input voltage from a load cell at span. Unit : mV/V.
This value and the value of
in span calibration during the calibration with an
actual load.
The calibration weight value corresponding to the
Input voltage at span of CAL18. When performing
digital span, CAL17,
required. The decimal point position is the same as
span
the setting of CAL02.
Input voltage
CAL17
0
Gravity acceleration of the place of calibration.
Unit : in m/s
Gravity acceleration of the place where the scale is
being used. Unit : in m/s
0 : Disables.
1 : Enables.
Description
CAL18
CAL19
Displayed weight
CAL18
and
CAL19
CAL18
and CAL19, "zero calibration" and "span
2
.
2
.
- 27 -
CAL19
are determined
and
CAL19
are
CAL18
in the "Setting List" at the end of the
Default
value
0
0.0000
3.2000
32000
9.8000
9.8000
0