Chapter 5 To Trigger the Oscilloscope
B
Slope time
When the Slope Condition is set to
press Time and use
ns to 10 s.
When the Slope Condition is set to
and Lower Limit and use
range of time upper limit is from 16 ns to 10 s. The range of the time lower limit
is from 8 ns to 9.99 s. Note that the time lower limit must be lower than the
upper limit.

Vertical Window:

Press Vertical to select the desired vertical window. Note that under the "Slope"
trigger menu, you can also press down the trigger level knob continuously to switch
the vertical window.
You can select the boundaries of the trigger level and then use TRIGGER
to adjust the trigger level. During the adjustment, two orange trigger level lines and
two trigger marks (
the rotation of the knob, while at the same time, the trigger level value and the slope
value are displayed at the lower left corner of the screen. When stopping turning the
knob, the trigger level lines and trigger marks disappear in about 2 s. When the
Slope Condition is set to
range will be displayed at the lower left corner of the screen.
UpLevel LowLevel
=
SlewRate
Time
DS1000Z User's Guide
A
Positive
to input the desired value. The range available is from 8
and
) appear on the screen and move up and down with
The upper limit of trigger level (Up Level)
The lower limit of trigger level (Low Level)
,
or
to input the desired values respectively. The
or
, the current trigger level and slope
UpLevel LowLevel
=
MinRate
UpLevel LowLevel
=
MaxRate
,
or
, press Upper Limit
MaxTime
MinTime
RIGOL
,
LEVEL
5-13