Rigol DS1000Z Series User Manual page 250

Digital oscilloscope
Hide thumbs Also See for DS1000Z Series:
Table of Contents

Advertisement

RIGOL
Horizontal
Time Base Scale
Time Base
[1]
Accuracy
Time Base Drift
Max Delay Range
Time Base Mode
Number of X-Ys
Waveform Capture
[2]
Rate
Vertical
Bandwidth (-3dB)
Single Bandwidth
Vertical Resolution
Vertical Scale
Offset Range
[1]
Bandwidth Limit
Low Frequency
Response
(AC coupling,
-3dB)
[1]
Rise Time
DC Gain
17-2
1000 Vpk
With RP2200 10:1 probe: CAT II 300 Vrms
5 ns/div to 50 s/div
≤ ± 25 ppm
≤ ± 5 ppm/year
Pre-trigger (negative delay): ≥1/2 screen width
Post-trigger (positive delay): 1 s to 5,000 s
Y-T, X-Y, Roll, Delayed
1 path
30,000 wfms/s (dots display)
DS1104Z: DC to 100 MHz
DS1074Z: DC to 70 MHz
DS1104Z: DC to 100 MHz
DS1074Z: DC to 70 MHz
8 bit
1 mV/div to 10 V/div
1 mV/div to 499 mV/div: ± 2 V
500 mV/div to 10 V/div: ± 100 V
20 MHz
≤5 Hz (on BNC)
DS1104Z: 3.5 ns
DS1074Z: 5 ns
<10 mV:±4% full scale