Press Storage  Default to restore the instrument to its default configuration.
1.
2.
Connect the ground alligator clip of the probe to the "Ground Terminal" under
the probe compensation signal output terminal.
3.
Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe.
Press AUTO.
4.
5.
Observe the waveform on the display. In normal condition, the display should be
a square waveform as shown in the figure below:
6.
Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform
Compensation" in the next section.
WARNING
To avoid electric shock during the use of probe, please make sure that
the insulated wire of the probe is in good condition and do not touch
the metallic part of the probe when the probe is connected to high
voltage source.
1-8
Figure 0-6 To Use the Compensation Signal
Figure 0-7 Square Waveform
Chapter 1 Quick Start
Compensation Signal Output Terminal
Ground Terminal
DS1000Z User's Guide