Using the Oscilloscope

To access the channel setup

• Click the channel number button at the top of the waveform display.
• Select the desired channel from the Setup menu.
Click one of the
channel buttons at the
top of the waveform
display...
Figure 34 Accessing the Channel Setup
66
or
With the channel setup dialog box, you can set the vertical scaling,
offset, input coupling, and input impedance. It also lets you specify the
characteristics of the probe, using the Probes button. You can set
attenuation ratio, attenuation units, and measurement units for the
probe, or perform a probe calibration.
For Agilent Technologies probes that are compatible with AutoProbe
interfaces, the oscilloscope will automatically set these characteristics
(except for skew) after identifying the probe when it is connected to the
channel input.
...or select a
channel from
the Setup
menu...
Agilent Infiniium 90000 Q-Series Oscilloscopes User's Guide
...to access the channel
setup dialog box.