To connect threaded RF cables

You can connect threaded RF cables to the Infiniium 90000 Q- Series
oscilloscope inputs.
For input channels 1 through 4, you can connect 3.5 mm threaded RF
cables, and you can use the automatic torque "clutch" mechanism to
ensure a consistent 8 in. lbs. connection.
The Aux Out and Cal Out connectors also work with 3.5 mm threaded RF
cables.
For RealEdge input channels 1R and 3R (on the 50 GHz and 62 GHz
bandwidth models), you can connect 1.85 mm threaded RF cables, and you
will need a torque wrench to apply the necessary 8 in. lbs. of torque.

To tilt the oscilloscope for easier viewing

There are tabs under the oscilloscope's front feet that can be flipped out
to tilt the oscilloscope.
Figure 5
Agilent Infiniium 90000 Q-Series Oscilloscopes User's Guide
Flip-Out Tabs Under Front Feet
Setting Up the Oscilloscope
1
21