Section 4 Startup

Complete the steps that follow each time the instrument connects to the power source. Make sure
that all of the electrical and plumbing connections are installed before startup.
1. Open the sample line to let sample flow through the analyzer.
2. Examine the plumbing for fluid leaks. Stop the leaks if found.
3. Make sure that the air circuit breaker switch is in the on position. Refer to
4. Push to close the analysis section. Listen to make sure that the panel moves fully into the latch.
5. Close the doors.
6. Supply power to the analyzer through a grounded electrical receptacle. Refer to
The display light comes on. When power is applied to the instrument for the first time, the
instrument shows the initialization guide. Refer to