8.14 Replace the piston pump
Replace the piston pump after 24 months of operation.
1. Push the circular START/STOP button to stop the analyzer operation. Select YES to confirm. The
analyzer starts a sequence of procedures to prepare the system for maintenance. Wait for all of
the procedures to stop.
2. Disconnect power from the instrument.
3. Remove the safety panel.
4. Replace the piston pump. Refer to the illustrated steps that follow.
5. Install the safety panel.
6. Push the fluids panel closed. Listen to make sure that the panel moves fully into the latch.
7. Close the doors.
8. Push the circular START/STOP button to start the analyzer operation. Select YES to confirm. The
analyzer starts a series of procedures to prepare the system for operation.
English 47