  • Page 2 客服:400 688 9809 26197-18 BODTrak™ INSTRUMENT MANUAL 网址:www.mamots.com 电话:0431-85178766 © Hach Company, 1995 - 1998. All rights reserved. Printed in the U.S.A. ar/lr 12-8-95 1ed Rev. 5, 7/98...
  • Page 3 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 4 The Model BODTrak has been tested and is certified as indicated to the following instrumentation standards: FCC Part 15, Sub-Part B, Class A Limits: Supporting...
  • Page 5 • Disconnect power from the BODTrak instrument to verify the instrument is the source of the interference. • If the BODTrak is plugged into the same outlet as the device with which it is interfering, try another outlet. • Move the BODTrak away from the device receiving the interference.
  • Page 6: Table Of Contents

    SPECIFICATIONS ..............vii SAFETY ..................ix OPERATION ................1 SECTION 1 INTRODUCTION ..........3 1.1 Before Operating the BODTrak Instrument ....3 1.2 Unpacking the BODTrak Instrument ......3 1.3 Applying Power.............. 4 1.4 The BOD Test ..............4 SECTION 2 INSTRUMENT OPERATION ......7 2.1 Principles of Operation...........
  • Page 7 4.5.3 Seed Acclimatization ........... 35 4.6 pH Effect ..............36 4.7 Supersaturation............. 36 SECTION 5 MAINTENANCE ..........37 5.1 Cleaning the BODTrak Instrument ......37 5.1.1 Sample Bottles ............. 37 5.1.2 Stir Bars and Seal Cups........37 SECTION 6 TROUBLESHOOTING........39 6.1 Adjusting the Calibration ..........
  • Page 8: Specifications

    客服:400 688 9809 SPECIFICATIONS Range: Selectable, 0-35, 0-70, 0-350, and 0-700 mg/L Dimensions: 33 x 30.5 x 12.7 cm (13 x 12 x 5 inches) External Power Supply: Input–115/230 VAC, 50/60 Hz Output–20.5 VRMS UL, CSA, and TUV approved Capacity: Six 473-mL bottles Shipping weight: 9 pounds Operating temperature: 20 °C (68 °F) Storage temperature: Below 40 °C (104 °F)
  • Page 9 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766 viii...
  • Page 10: Safety

    客服:400 688 9809 SAFETY Before attempting to unpack, set up, or operate this instrument, please read this entire manual. Pay particular attention to all warnings, cautions and notes. Failure to do so could result in serious injury to the operator or damage to the equipment. Use of Hazard Information If multiple hazards exist, the signal word corresponding to the greatest hazard shall be used.
  • Page 11 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 12: Operation

    WARNUNG Da das Arbeiten mit chemikalischen Proben, Standards, Reagenzien und Abfällen mit Gefahren verbunden ist, empfiehlt die Hach Company dem Benutzer dieser Produkte dringend, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien oder Biogefahrgut vertraut zu machen und alle entsprechenden Materialsicherheitsdatenblätter aufmerksam zu lesen.
  • Page 13 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 14: Section 1 Introduction

    • Power supply – switchable between 115V or 230V • Stopcock grease • One package Nutrient Buffer Solution Pillows • One package Lithium Hydroxide Powder Pillows • Power cords, 115 and 230 V If any items are missing or damaged, please contact the Hach 网址:www.mamots.com 电话:0431-85178766...
  • Page 15: Applying Power

    A UL/CSA approved 115 Vac power cord with a NEMA 5-15P style plug and a 230 Vac Harmonized power cord with a Continental European plug are supplied with the BODTrak Instrument. The plug supplies AC power to the Power Supply IEC Universal Connector (see Figure 1).
  • Page 16 This allows operators to estimate plant operating efficiency and determine proper treatment procedures. The BODTrak method is an easy and direct method of measuring BOD. It makes complex chemical analysis unnecessary because the BODTrak instrument takes a physical measurement of the oxygen consumed by a test sample.
  • Page 17 客服:400 688 9809 • The BODTrak method continuously removes carbon dioxide ) from the system so that the pressure difference observed is proportional to the amount of oxygen used. • Bubbles formed by degassing can cause errors when samples are warmed to the experimental temperature. The BODTrak compensates for this occurrence by starting the test after the equilibrium temperature has been reached.
  • Page 18: Section 2 Instrument Operation

    Stirring helps transfer oxygen from the air to the sample and helps simulate natural conditions. 2.1.2 Pressure Sensor Function The BODTrak is sealed to prevent external atmospheric pressure changes from affecting the BOD reading. Pressure sensors monitor air pressure within the sample bottles, and when the air pressure drops, the pressure change is converted to mg/L BOD.
  • Page 19: Key Functions

    客服:400 688 9809 2.2 Key Functions The BODTrak control panel (see Figure 2) includes the channel selector, located centrally on the panel; the display screen, to the right of the channel selector; and the control keys, located below the display screen.
  • Page 20 The cursor is automatically placed at the most recently collected data point whenever a channel is displayed. To start a test, follow the procedure in section 3.1 BODTrak Test Procedure and proceed as follows: 1. Connect the tube cap for the measurement channel to the bottle being tested, and firmly tighten the cap.
  • Page 21: Setting The Clock

    客服:400 688 9809 The BOD test automatically ends after the selected test duration (5.25, 7, or 10.50 days), if not ended manually before then. The status annunciator in the lower right of the display will change from indicating test completion. To end a test END, manually, proceed as follows: 1.
  • Page 22: Rs232 Interface

    MODE COMn, 9600,N,8,1 Where n is the port number 1 or 2. The BODTrak instrument is wired as Data Communication Equipment (DCE) and operates at 9600 baud with 8 data bits, no parity, and one stop bit. The pinouts of the necessary cables are shown in Table 3 and Table 3.
  • Page 23: Commands

    Computer 9-pin Connector, Socket Connector, Socket Signal Name Signal Name Not Used Not Used Table 3: Wiring Cable from BODTrak to Serial Printer BODTrak 9-pin Printer 25-pin Connector, Socket Connector, Plug Signal Name Signal Name * Epson LQ-570+ (Cat. No. 48236-00) 2.4.2 Commands...
  • Page 24 客服:400 688 9809 ASCII characters followed by a carriage return. Each command (such as selecting a channel) performs a single action as soon as the instrument receives it. The instrument then sends a carriage return and line feed to indicate completion. Query commands, such as Sx, GC, and GA respond with additional characters before the final carriage return and line feed.
  • Page 25 客服:400 688 9809 Command Summary Table 4 summarizes the RS232 commands. Notations x, xx , or xxx denote the maximum number of places for a numeric parameter. Table 4: Command Summary Command Function Initializes all date/time registers to their current values and returns the current values.
  • Page 26 客服:400 688 9809 Table 4: Command Summary (cont.) Command Function Sets dilution concentration. Functions only when the display shows the ON screen. For example, sending the S#xxx S#700, S#350, or S#70 parameter sets a 0-700, 0-350, or 0-70 mg/L test range respectively. Any other number results in a 0-35 mg/L test range.
  • Page 27: Downloading And Printing Test Results

    2.4.3 Downloading and Printing Test Results Downloading Test Results Transfer the test results from the BODTrak to the computer by sending the appropriate set channel command to select the channel to be downloaded. For example, if you want to download the data from channel 1, type S1 followed by a carriage return.
  • Page 28 客服:400 688 9809 Figure 3 shows channel number, start date, start time, and the format of the downloaded data. BOD values in mg/L follow. Only the first data points, of a maximum of 480 evenly spaced points, are shown in this example. Each line ends with a carriage return and a line feed.
  • Page 29 客服:400 688 9809 Printing Test Results Transfer the test results from the BODTrak to the Epson LQ-570+ printer by pressing and holding the appropriate channel number for several seconds. The instrument will send a copy of the graphical display as well as a truncated data stream (127 data points).
  • Page 30: Section 3 The Bod Test

    客服:400 688 9809 SECTION 3 THE BOD TEST 3.1 BODTrak Test Procedure WARNING Follow proper laboratory technique when handling materials, samples, and waste. Wear eye protection and latex gloves to avoid illness or infection. ADVERTÊNCIA Siga as tecnicas proprias de laboratorio quando manusear materiais, amostras, e descartes.
  • Page 31: Bodtrak Test Procedure

    客服:400 688 9809 BODTrak Test Procedure, continued Heat or cool the Using a clean Place a 3.8-cm sample to within 2 °C graduated cylinder, (1½-inch) magnetic stir of its incubation pour the correct bar in each temperature (typically sample volume into a sample bottle.
  • Page 32 客服:400 688 9809 BODTrak Test Procedure, continued Add the contents Apply Stopcock Place a seal cup in of one BOD Nutrient Grease to the seal lip of the neck of each bottle. Buffer Pillow to each each bottle and to the bottle for optimum top of each seal cup.
  • Page 33 客服:400 688 9809 BODTrak Test Procedure, continued CAUTION Lithium Hydroxide is strongly alkaline and irritating to the skin and eyes. Practice safe laboratory techniques. PRECAUÇÃO O hidróxido de lítio é fortemente alcalino, caústico e irrita a pele e os olhos. Use praticas seguras de laboratorio.
  • Page 34 客服:400 688 9809 BODTrak Test Procedure, continued Start the instrument Make sure all stir To select a (connect the electrical bars are rotating. If a test duration, plug and turn the stir bar slides to the simultaneously press < instrument on).
  • Page 35 客服:400 688 9809 BODTrak Test Procedure, continued To start a test, Press the key. For 0–350 mg/L > press the channel A menu for selecting range, press the number corresponding the BOD range will (right) key. For 0–700 > to the bottle.
  • Page 36 客服:400 688 9809 BODTrak Test Procedure, continued Press and hold the Read the BOD Use a brush and key to start a test. results directly from hot soapy water to A graph will be the BODTrak display clean all bottles, stir displayed.
  • Page 37: Interpreting Test Results

    客服:400 688 9809 3.2 Interpreting Test Results If the test proceeds correctly, the display should produce a curve similar to Curve A in Figure 4. Figure 4 Example of BOD Curves If such a curve does not occur, one or more of the following problems may have occurred: bottle leak, time lag, high oxygen demand, or nitrification.
  • Page 38: Time Lag

    Control nitrification problems with Hach Nitrification Inhibitor (Cat. No. 2533-35). Dispense the inhibitor powder into an empty sample bottle and then add the sample. When using the Hach Dispenser cap (Cat No. 459-01), dispense two shots (approximately 0.16 grams) into the empty bottle.
  • Page 39: Accuracy Check

    客服:400 688 9809 3.3 Accuracy Check To check test accuracy and proper performance of the BODTrak instrument, test a standard BOD sample and evaluate it for one or more of the chemical, physical, or biological effects described in 3.2 Interpreting Test Results.
  • Page 40 Follow the same procedure used to determine the BOD of a sample. See the iconed procedure, section 3.1 BODTrak Test Procedure for details. 7. To determine the BOD result, see section 4.3.2 Determining BOD of Sample. The corrected BOD of the standard solution should be 198 ±30.5 mg/L.
  • Page 41 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 42: Section 4 Special Considerations

    客服:400 688 9809 SECTION 4 SPECIAL CONSIDERATIONS 4.1 Range and Volume Selection Because the BODTrak instrument can measure various waste substances, a variety of total oxygen demand test results can be obtained. The instrument provides four direct-reading ranges (0-35, 0-70, 0-350, and 0-700 mg/L). Use Table 5 on page 20 to select the volume and range corresponding to the expected sample BOD.
  • Page 43: Preparing Several Identical Samples

    客服:400 688 9809 4.2.2 Preparing Several Identical Samples Perform a single dilution for all samples when several identical samples are needed. After the dilution, multiply the reading by the dilution factor. Example: Prepare a 1:5 dilution by multiplying the original sample volume by 5 and adding dilution water until the new volume is obtained.
  • Page 44: Determining Bod Of Sample

    客服:400 688 9809 4.3.2 Determining BOD of Sample After determining the BOD of the seed, apply the following formula to determine the sample’s BOD. BOD observed Decimal fraction of seed used x BOD seed – BOD sample ----------------------------------------------------------------------------------------------------------------------------------------------------------------------- - Decimal fraction of sample used Example: A seeded sample is 10% seed and 90% sample (by volume).
  • Page 45: Sample Temperature

    The American Public Health Association (APHA) recommends a solution temperature of 20 ±1 °C (68 °F) for conducting the BOD test. Obtain this temperature by placing the BODTrak instrument in an appropriate incubator and adjusting the temperature until the solution reaches 20 ±1 °C. An undercounter BOD Incubator and a combination BOD Incubator/Refrigerator are available (see PARTS &...
  • Page 46: Other Toxic Materials

    NOTE: To determine the concentrations of these and other materials, see the Hach “Water Analysis Handbook.” Dilute the sample with distilled water to eliminate the effect of these materials. The correct BOD is obtained when two successive dilutions result in the same sample BOD value.
  • Page 47: Ph Effect

    客服:400 688 9809 For more information, Standard Methods for the Examination of Water and Wastewater, 18th edition emphasizes the importance of selecting the proper seed for specific wastes. If the waste sample to be tested contains toxic materials such as phenol, formaldehyde, or other microbic inhibitory agents, use acclimated seed.
  • Page 48 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 49 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 50: Section 5 Maintenance

    客服:400 688 9809 SECTION 5 MAINTENANCE 5.1 Cleaning the BODTrak Instrument Use the following methods to clean the BODTrak instrument after each test to assure accuracy of future BOD tests. 5.1.1 Sample Bottles 1. After each test, empty the sample bottles and rinse them several times with hot water.
  • Page 51 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 52: Section 6 Troubleshooting

    See section 6.1 Adjusting the values Calibration. Fan not working Discontinue use. Contact the Hach Service Department. 6.1 Adjusting the Calibration If the instrument records consistently high or low values for the accuracy check in Section 3.3, it may be necessary to adjust the calibration as follows: 1.
  • Page 53 客服:400 688 9809 4. Record all of the factory installed calibration values below before proceeding! If these values are not recorded at this time they will be lost. Channel 1________ Channel 2________ Channel 3________ Channel 4________ Channel 5________ Channel 6________ 5.
  • Page 54: General Information

    客服:400 688 9809 GENERAL INFORMATION At Hach Company, customer service is an important part of every product we make. With that in mind, we have compiled the following information for your convenience. 网址:www.mamots.com 电话:0431-85178766...
  • Page 55 客服:400 688 9809 网址:www.mamots.com 电话:0431-85178766...
  • Page 56: Parts & Accessories

    Incubator/Refrigerator, Model 207, 120V......... 26198-00 Incubator/Refrigerator, Model 207, 240V......... 26198-02 IncuTrol/2, Temperature Regulator, 115V ........2597-00 IncuTrol/2, Temperature Regulator, 220V ........2597-02 Manual, BODTrak Instrument ..........26197-18 Pipet filler .................. 12189-00 Pipet, serological, glass 1.0 mL..................532-35 10 mL..................532-38 Power cord, 18/3 SVT 7.5’, 10A-125 Vac...
  • Page 57 客服:400 688 9809 PARTS & ACCESSORIES, continued Sampler, sewage ................427-00 Seal Cup ..................10977-52 Software, Data Capture Package ..........49665-00 Stir Bar, magnetic ..............10764-16 Support Stand, buret ..............329-00 Thermometer, -20 to 110 °C............566-01 Water Still, 240V............... 14867-00 REAGENTS Description Cat.
  • Page 58: How To Order

    Loveland, Colorado 80539 U.S.A. By Telex: 160840 (Hach Loveland) By FAX: 970-669-2932 (Hach Loveland) Information Required • Hach account number (if available) • Billing address • Shipping address • Your name and phone number • Purchase order number • Catalog number •...
  • Page 59 HOW TO ORDER, continued Call or write for free quotations. Outside the United States Hach maintains a network of dealers and distributors throughout the world. In Europe, Mediterranean Africa, and the Middle East, customers can contact their local distributor or obtain help from: Hach Europe, S.A./N.V.
  • Page 60: Repair Service

    客服:400 688 9809 REPAIR SERVICE Authorization must be obtained from Hach Company before sending any item for repair. Please contact the Hach Service Center serving your location. In the United States: Hach Company 100 Dayton Ave. Ames, Iowa 50010 800-227-4224 (USA only)
  • Page 61: Warranty

    FAX: (970) 669-2932 FAX: (32) (81) 44.13.00 Telex: 160840 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING In the U.S.A. - Call 800-227-4224 toll-free for more information. Outside the U.S.A. - Contact the Hach office or distributor serving you. 网址:www.mamots.com 电话:0431-85178766...
  • Page 62: Index

    客服:400 688 9809 INDEX control panel 8 current time 14 accuracy check 28 cursor 9 air pressure 7 curve 26 alkaline permanganate 31 data bits 11 bacteria 4, 27, 31 data flow 16 baud 11 data format 16 Biochemical Oxygen Demand data transmission 15 (BOD) 4 date 14...
  • Page 63 客服:400 688 9809 INDEX, continued specifications vii, 3 standard sample 28 manually stop test 10 start time 9 minute 14 status annunciator 9, 10 month 14 stir bars 23 stirring 5 natural conditions 5 stored data 11 nitrification 27 nitrogen 27 test data 15 nutrient buffer solution 29 test duration 10, 23...