3.1 Installation guidelines

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Fire hazard. This product is not designed for use with flammable liquids.
• Install the analyzer indoors, in a non-hazardous environment.
• Install the analyzer in a clean, dry, well-ventilated and temperature-controlled location.
• Install the analyzer as near to the sampling point as possible.
• Do not install the analyzer in direct sunlight or near a heat source.
• Make sure that there is sufficient clearance to make plumbing and electrical connections.
• Make sure to leave sufficient space in front of the analyzer to open the analyzer door.
• Make sure that the ambient conditions are within operating specifications. Refer to
on page 3.

3.2 Mechanical installation

3.2.1 Attach the instrument to a wall

Personal injury hazard. Make sure that the wall mounting is able to hold 4 times the
weight of the equipment.
Personal injury hazard. Instruments or components are heavy. Use assistance to install
or move.
This instrument is rated for an altitude of 2000 m (6562 ft) maximum. Use of this instrument at an
altitude higher than 2000 m can slightly increase the potential for the electrical insulation to break
down, which can result in an electric shock hazard. The manufacturer recommends that users with
concerns contact technical support.