CORRECTIVE MAINTENANCE

DIGITAL READOUT MESSAGE CODES

Listed below are message codes which may be displayed on the Digital Readout if a mistake is made by the
operator (e.g., wrong password entry) or if the system has malfunctioned.
DISPLAY MESSAGE
The PACE Service department can be reached at:
Table 1. Digital Readout Message Codes
78
DESCRIPTION
The incorrect Password has been entered. The
displayed message will time out after 3 seconds and
revert to normal operation. Enter the correct Password.
Circuitry within the ThermoFlo prevents the handpiece
from overheating. If this message is displayed, allow
the handpiece to cool before resuming operation.
The handpiece heater assembly sensor is open.
Replace heater assembly.
The ThermoFlo has detected a ground fault in the
handpiece. Check the GFI Light on the rear panel. If
the light is out, turn power off and then back on. If the
GFI light is still out, a ground fault has been detected.
Contact PACE Customer Service for assistance.
The power source blower unit is not running. Contact
the PACE Service Department for assistance.
Tel. (301) 490-9860
FAX (301) 483-7030