8.6.1.1.
8.6.1.2.
LACO Technologies
DANGER: Life threatening hazard from electric shock
Considerable voltages arise inside the device. Touching parts where electrical
voltage is applied can result in death.
• Disconnect the device from the power supply before any cleaning or
maintenance. Ensure the electric power supply is reconnected with
authorization.
Fuses
Loc. ID Designation
1
Mains Fuse
2
24 V Fuse
Replacing the main fuse
•
Switch the device off at the mains plug.
•
Disconnect the device from the power supply and pull out the main plug.
•
Pry the cover of the fuse holder and fold it to the side.
•
Remove the two fuse holders and replace the defective fuses (10.0 AT, 250 V,
Ø5 x 20 mm).
•
Lock the fuse holders back into place.
•
Close the cover.
Replacing terminal block fuses
•
Turn off power to unit.
•
Open enclosure.
•
Locate fuse holder terminal bock and use hand to swing to open position.
•
Open fuse terminal block fuse compartment to reveal defective use and
replace (3.0 AT, 250 V, Ø5 x 20mm).
ATLAS Operations and Maintenance Manual
8 MAINTENANCE AND SPARE PARTS
Rating
Quantity Part Number
T 10.0 A 2
LMSA3844-C
T 3.0 A
1
LMSA3851
59