LACO Technologies ATLAS Operation And Maintenance Manual page 55

Leak test controller
Table of Contents

Advertisement

LACO Technologies
2. Press green start button to start the test cycle.
3. The controller will begin stepping through the part charging sequence
(Evac, He, Fill, etc. depending on configuration). Once the helium fill is
complete, the message below will appear.
4. When the operator presses the Ok button the leak test will begin. At this
point the operator can begin sniffing the part to be tested. During the sniff
test, the ATLAS will store and display the peak leak rate value that is
observed. To clear this value at any time during the test the operator can
press the Clear Peak Value icon at the bottom left of the test screen.
5. When the operator is finished with the sniff test, they must acknowledge it
by pressing the OK button on the right side of the screen.
ATLAS Operations and Maintenance Manual
6 OPERATION
47