7 TROUBLESHOOTING
56
Fault Description
LD Failed to Start Cal
Calibration Failed
ATLAS Operations and Maintenance Manual
Fault Causes
The ATLAS was unable to
initiate a calibration cycle
in the Titan Versa.
The measured leak rate is
outside the designated
limits of where it is
expected to read.
Fault Troubleshooting
•
Check communication
cables between ATLAS
and Titan Versa.
•
Confirm that the part
in the chamber is
properly connected
and equipped with the
Calibrated Leak
Standard that is
entered on the Cal
Settings screen.
•
Change the Min
and/or Max Cal Leak
Factors on the Cal
Settings Screen.
•
Verify the Calibrated
Leak Value entered on
the Cal Settings
screen.
•
Test the Calibrated
Leak functionality in
manual mode to
ensure the leak is
reading match the
stated values.