LACO Technologies ATLAS Operation And Maintenance Manual page 64

Leak test controller
Table of Contents

Advertisement

7 TROUBLESHOOTING
56
Fault Description
LD Failed to Start Cal
Calibration Failed
ATLAS Operations and Maintenance Manual
Fault Causes
The ATLAS was unable to
initiate a calibration cycle
in the Titan Versa.
The measured leak rate is
outside the designated
limits of where it is
expected to read.
Fault Troubleshooting
Check communication
cables between ATLAS
and Titan Versa.
Confirm that the part
in the chamber is
properly connected
and equipped with the
Calibrated Leak
Standard that is
entered on the Cal
Settings screen.
Change the Min
and/or Max Cal Leak
Factors on the Cal
Settings Screen.
Verify the Calibrated
Leak Value entered on
the Cal Settings
screen.
Test the Calibrated
Leak functionality in
manual mode to
ensure the leak is
reading match the
stated values.