Leak test controller
Table of Contents

Advertisement

Calibrated Leak Value
Hours Cal Valid
Cycles Cal Valid
Cycles Cal Valid
Cal Factor
QCal Final
Cal Factor
LACO Technologies
The value listed on the calibration certificate of the calibrated leak used for
calibrating the system.
When a successful calibration is completed the timer is reset. After the timer
expires the system will require a calibration.
When a successful calibration is completed the counter is reset. After the counter
expires the system will require a calibration. NOTE: if both the Cycles/Hours Cal
Valid parameters are set to 0, the system will never require a calibration.
This number creates a window that the calibrated leak must be between at the end
of a calibration to be a good calibration.
This value is calculated by taking the calibrated leak value / raw leak rate reading
at the end of the calibration cycle.
The leak rate reading using the default Cal factor at the end of a calibration.
Resets the Cal Factor to the Default Cal Factor.
ATLAS Operations and Maintenance Manual
6 OPERATION
41