LACO Technologies ATLAS Operation And Maintenance Manual page 33

Leak test controller
Table of Contents

Advertisement

Evac Pump Start Delay
Ticket Printer
6.5.2.2.
Select Process
Enabled
Start Delay
Stop Delay
Settle Time
LACO Technologies
Venturi
Always On
Powered As Needed
Delay time from when the pump is powered on to when the process continues.
If a printer is included, ensure that this option is checked and that it is attached
with the included USB to RS232 adapter. This cannot be on if communication is
configured to RS232.
Part Charging Settings
Select the process to edit. The four available options are:
Helium
Evac
Air
Vent
Check the box to show the parameters of the selected process. Unchecking the
box will also disable the process valve in manual mode.
Delay time to allow the valve to fully open after the output is turned on.
Delay time to allow the valve to fully close after the output is turned off.
After the pressure has reached the setpoint, the settle delay timer allows for the
pressure to equalize before evaluating if the pressure is still in tolerance. This settle
timer applies to both fill and evac operations.
ATLAS Operations and Maintenance Manual
6 OPERATION
25