Upper/Lower Limit Of Signal Output Range - Optex SE-pH-D Instruction Manual

Ph/orp sensor
Table of Contents

Advertisement

Upper/lower limit of signal output range

For the item by 4-20 mA signal output, set measurement value to output
at lower limit (Lo, 4 mA) and upper limit (Hi, 20 mA).
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "6.3 4-20 mA Setting", " ◆ Signal output range lower limit"
and " ◆ Signal output range upper limit".
Output item
[Display]
pH
[pH]
Temperature
[Temp.]
pH
Electromotive
force
[pH mV]
ORP
ORP
[ORP]
The setting unit is based on the display resolution of the item set for the
display item. (For display resolution, refer to "15 Specifications" P.54)
22
20 mA
Signal output
4 mA
Measurement
value at 4 mA
Factory default
setting
Lower limit: 0.00
Upper limit: 14.00
Lower limit: 0.0
Upper limit: 60.0
Lower limit:
-500.00
Upper limit:
500.00
Lower limit:-1500
Upper limit:1500
Measurement
value at 20 mA
Signal output
range
Lower limit to
Upper limit
0.00 to 14.00
1.00 pH
0.0 to 60.0
5.0°C
-500.00 to
2.00 mV
500.00
-1500 to 1500 100 mV
Minimum
setting
range