Cleaning Interval (Relay) - Optex SE-pH-D Instruction Manual

Ph/orp sensor
Table of Contents

Advertisement

Select [Manual Cleaning] using
3
The cleaning device is activated
and the display returns to the
measurement value display.
The sensor under cleaning shows
"CLEAN" on the measurement
value display.
In addition, "C" is displayed in the
sensor selection display.

Cleaning interval (relay)

Set a time interval to activate the cleaning device.
The settings are to activate the cleaning device using a relay output
("Cleaning output setting" P.24).
Cleaning
activation
Cleaning
Cleaning
device
Measurement
Measurement
value hold
value
Select an item from the following.
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "9 Cleaning" and " ◆ Cleaning Interval (Relay)".
Item
[Display]
Cleaning Interval
(Relay)
[Cleaning Interval-RLY]
46
, and press
Cleaning interval (relay)
Post-
cleaning
standby
Normal measurement
OFF (No cleaning)
10 to 1440 minutes
(by 10 minutes)
.
Cleaning interval (relay)
Post-
Cleaning
cleaning
activation
standby
Cleaning
Measurement
value hold
Setting range
Normal measurement
Factory
default setting
30 minutes