1-Point Calibration (Calibration Mode: Manual) - Optex SE-pH-D Instruction Manual

Ph/orp sensor
Table of Contents

Advertisement

1-point calibration (Calibration mode: Manual)

Perform calibration with the pH standard solution set in step 4 of "Calibration
standard selection" P.18.
・ The 1-point calibration is for pH measurement. It does not affect
ORP measurement.
Caution
・ In 1-point calibration, span errors can be minimized by using
a calibration solution (standard solution) that is close to the
expected pH value of the measurement water.
Clean the sensor body and electrode surface.
1
Prepare the calibration solution (standard solution) of the selected
2
pH for the 1st calibration solution (standard solution) and immerse
the sensor.
Allow the sensor to settle to the temperature of the calibration
3
solution (standard solution) for about 5 minutes with little change
in temperature, then check that there are no bubbles on the
electrode surface.
Press
4
The sensor selection display is displayed.
Select a channel connected to
5
SE-pH-D using
The calibration menu is displayed.
Select [1-Point Calibration] using
6
28
.