Upper/Lower Limit Of Signal Output Range - Optex DOS-20 Instruction Manual

Optical dissolved oxygen sensor
Table of Contents

Advertisement

Upper/lower limit of signal output range

For the item by 4-20 mA signal output, set measurement value to output
at lower limit (Lo, 4 mA) and upper limit (Hi, 20 mA).
Signal output
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.3 4-20 mA Setting", " ◆ Signal output range lower limit"
and " ◆ Signal output range upper limit".
Output item
[Display]
Dissolved oxygen
concentration [DO]
Temperature [Temp.] Lower limit: 0.0
Dissolved oxygen %
saturation [%Sat.]
Oxygen partial
pressure [O-Press]
The setting unit is based on the display resolution of the item set for the
display item. (For display resolution, refer to "15 Specifications" P.52)
16
20 mA
4 mA
Measurement
value at 4 mA
Factory default
setting
Lower limit: 0.00
Upper limit: 50.00
Upper limit: 50.0
Lower limit: 0.0
Upper limit: 200.0
Lower limit: 0.0
Upper limit: 150.0
Measurement
value at 20 mA
Signal output
Minimum setting
range
Lower limit to
Upper limit
0.00 to 50.00 2.00
0.0 to 50.0
10.0
0.0 to 200.0
10.0
0.0 to 150.0
10.0
range
Upper limit -
Lower limit ≥

Advertisement