8.
Press the Digital Interface "Output" key of the 1053. The data will be recorded by
the 7400, which will stop automatically once the data has been transferred.
9.
Note down the File No. and the set-up characteristics for reference purposes.
The Digital Cassette Recorder can also be used to record dynamic-data read out during
a test run. In this case omit steps 8 and 9, and follow the procedure given in section
4.2.2, steps 4 or 5.
To set the test set-up parameters, by recalling data from the 7400, proceed as follows:
1.
Follow steps 1 to 4 listed above.
2.
Select the front-panel test set-up to be set by the 7 400.
3.
Press the Digital Interface "Input" key.
4.
Locate the correct file (refer to the instruction manual for the 7400).
5.
Press the Read File/Block "One F/S" key. The selected front-panel test parameters
of the 1 053 will be set to the recorded values.
4.2.5. Graphics Recorder Type 2313
32
11111111:~
······==
iii'
li
llill
i
Fig. 4.5. Graphics Recorder Type 2313
The Graphics Recorder Type 2313 can be used to print out a list of the test set-up
parameters, together with their set values. It can print out sets of the dynamic param-
eters: frequency, generator level (actual) and vibration-meter level.
To print out the dynamic parameters, during a test, proceed as follows:
1.
Connect the IEEE-488 Interface Bus (of the 1 053) to the IEC-625 Interface Bus (of
the 2313) using cable AO 0264.
2.
Set Address Switch 6 (of the 2313) to "0: System Controller not active".