5.4.29. Generator _Levei_Actual
Sets the generator level as in the front panel Generator "Level" function, when the
Generator Output is in "Operate" or "Operate -20dB" and the Compressor is not active.
Example of Use:
Generator _Levei_Actual
-35 . 2
- Sets the actual generator level to -35,2d8
Notes:
If Generator Output is in "Stand-by" or the Compressor is active when this message is
sent, an error will result.
If the Generator Output is in "Stand-by", the Generator_Levei_Programmed Message
can be used to set the level prior to switching On the generator output.
5.4.30. Generator _Levei_Actual?
Enables read out of the actual generator level.
Example of Use:
Generator _Levei_Actual?
enables read out of, e.g.:
G_L_A
-035. 2d8
Notes:
See section 5.4.29.
- If set using example in section 5.4.29
5.4.31. Generator _Levei_Programmed
50
Sets the generator level as in the front panel Generator "Level" function, when the
Generator Output is in "Stand-by".
Example of Use:
Generator _Levei_Programmed
-47. 5
Notes:
- Sets the programmed generator level
to
-47,5d8.
The programmed generator level can be set only if the 1053 is in Generator Output
"Standby" mode. If the Generator Output "Operate" is active
,
an error message will
result.
If the Generator Output is in "Operate", the Generator _Levei_Actual Message can be
used to change the actual level if the compressor is not active.