BRUEL & KJAER 1053 Instruction Manual page 65

Vibration exciter control
Table of Contents

Advertisement

Example of Use:
Readout_ Type
SINgle ,
<Mode)
General syntax for specifying a single read out
Readout_ Type
SINgle , 1
Specifies a single read out of frequency (mode 1 -see Notes)
Readout_ Type
Continuous,
<Interval)
,
<Mode)
General Syntax for specifying a continuous read out
Readout_ Type
Continuous , 2. 5, 5
Specifies continuous read out of frequency and meter level (mode 5 - see Notes), with
2,5 second intervals between read outs.
Readout_ Type
Start_On_Sweep ,
(Interval)
,
(Mode)
General syntax for specifying continuous read out, activated when the sweep is started
Readout_ Type
Start_On_Sweep , 1 . 25 , 6
Specifies continuous read out, activated when the sweep is started, of the Frequency,
Generator Level and Meter Level, with 1,25 second intervals between read outs
Notes:
If the parameter Start_On_Sweep is used, the read out will commence once the sweep is
activated and the Start_On_Sweep parameter will change to 'Continuous' read out.
The 'Interval' will be the interval between read-outs, entered in seconds. The range is 0,1
to 100 seconds in 0,01 s steps. When single read out is chosen, the 'Interval' parameter
must be omitted.
The 'Mode' will be a number from 1 to 6 corresponding to the interface Message(s)
listed in Table 5.3.
For continuous read-outs, the Sweep Count will be read out together with the param-
eters specified by the 'Mode' the first time, and each time the Sweep Count elapsed has
been increased.
Mode
Job Header
Section
1
Frequency
5.4.7
2
Generator _Levei_Actual
5.4.11
3
Meter _Level
5.4.29
4
Frequency, Generator _Levei_Actual
5
Frequency, Meter _Level
6
Frequency, Generator _Levei_Actual, Meter _Level
T01480GBO
Table 5.3. Interface Messages corresponding to the
'Mode'
number
of the Readout_ Type interface Message
57