Error; Externai_Filter - BRUEL & KJAER 1053 Instruction Manual

Vibration exciter control
Table of Contents

Advertisement

5.4.18. Error _Stop
Sets the 1053 to either stop or continue interface bus data transfer when an interface
error occurs.
Example of Use:
Error _Stop
No
or
Error _Stop
Yes
Notes:
- Default -
bus data transfer continues on detect-
ing an error. The interface message which causes
the error will be ignored until a message separator
('; ', LF or EOI with any character) is received
where message recognition will resume.
- Bus data transfer stops on detecting an error
The 1053 will halt data transfer by use of the NRFD (Not Ready For Data) handshake line.
This will be released by sending DEVICE CLEAR, manual reset 0, or power off then on.
5.4.19. Error _Stop?
Enables read out of the Error _Stop status.
Example of Use:
Error _Stop?
enables read out of:
E_S
No
or
E_S
Yes
Notes:
See Notes in section 5.4.18.
5.4.20. Externai_Filter
- For no error stop
- For stop on error
Turns on/off the compressor external filter and vibration meter external filter as in Spe-
cial Function 06 and 07 respectively.
Example of Use:
Externai_Filter
Cprs_ON , Meter _ON
- Turns On both the compressor external
filter and vibration meter external filter
Externai_Filter
Cprs_OFf , Meter _OFf
- Turns Off both the compressor external
filter and vibration meter external filter
5.4.21. Externai_Filter?
Enables read out of the compressor external filter and vibration meter external filter
status.
47