(b) Press Digital Interface "Output".
4.
For a continuous read-out of sets of the dynamic parameters synchronized with the
sweep start:
(a) Set the read-out time interval by pressing "x" (one or more numerals represent-
ing the time interval in seconds). The range is from 0,1 to 100 s with a resolution
of 0,01
s.
(b) Press Digital Interface "Output".
(c) Press Generator Output "Operate" and Sweep "Up" or "Down" to initiate a
frequency sweep and read-out of the dynamic parameters. The first set of dy-
namic parameters will coincide, in time, with the start of the sweep.
5.
For a continuous read-out of the dynamic parameters unsynchronized with the
sweep start:
(a) Press Generator Output "Operate".
(b) Press Sweep Control "Up" or "Down" to initiate the frequency sweep.
(c) Set the read-out time interval by pressing "x" (one or more numerals represent-
ing the time interval in seconds). The range is from 0,1 to 100 s with a resolution
of 0,01 s.
(d) Press Digital Interface "Output" to initiate read-out of the dynamic parameters
unaffected by the sweep start time.
These operational steps can be performed in the order: (a)-(c)-(d)-(b).
During a read-out, the Digital Interface "Output" LED will light to indicate that the 1053
is in Talk Only mode. When read-out is terminated, the LED will extinguish. Continuous
read-outs of dynamic parameters are terminated by pressing Digital Interface "Output"
for a second time.
4.2.3. Input of set-up data
30
AO 0264
Graphics Recorder
2313
AO 0231
·-
--~
~
--
~
..
.
131
:-·
--:
.
.
..
·
<i1ii@i
0087
ii
.•
Vibration Exciter
Power Amplifier
Vibration Exciter
Control 1053
2708
4802
+
Head 4818
870428
Fig. 4.3. Vibration-test system with Digital Recorders and interface
connections