3.
Press the Reset key (of the 2313); the Entering LED will light.
4.
Press the Exc./Ent key (of the 2313); the Listen LED will light.
5.
Initiate read-out using the procedure given in section 4.2.2 steps 4 or 5. This choice
depends on whether a synchronized or unsynchronized read-out is required.
6.
To terminate a print-out, press Digital "Interface Output" for a second time.
To print out a list of the front-panel test parameters, proceed as follows:
1.
Repeat steps 1 to 4, listed above.
2.
Set the Address Switch 2 (of the 1 053) to "0": IEEE Format, for an unformatted
print-out separated by semi-colons, or "1": Spec. Format, for a line-by-line print-
out.
3.
Perform Reset 0.
4.
Select the front-panel test set-up for print out by pressing Front Panel Set-up "No.".
Then press Digital Interface "Output" (refer to section 3.4.1 ).
Any single set-up parameter can be printed out in this way by pressing the associ-
ated key followed by Digital Interface "Output".
4.3. DIGITAL STROBOSCOPE TYPE 4913
Fig. 4.6. Digital Stroboscope 4913
The Digital Stroboscope Type 4913 can be used with the 1053 to provide slow motion or
frozen visual analysis of the structure under test (chapter 1, Fig. 1 0). The stroboscope
flash frequency is synchronized to the 1053 generator frequency. Consequently, an auto-
matic frequency sweep can be used for visual resonance detection and a manual fre-
quency sweep for detailed visual analysis at a resonance.