5.4.69. Sweep_Rate?
Enables read out of the sweep rate.
Example of Use:
Sweep_Rate?
enables read out of:
S_R
Log, 1 . 25E+00oct/min
- If set using first example in section 5.4.68
5.4.70. Tracking_Filter_Bandwidth
Sets the bandwidth of the tracking filter as in Special Function 05.
Example of Use:
Tracking_Filter _Bandwidth
10
- Sets the tracking filter bandwidth to 10%
Notes:
The possible settings are bandwidths of 3%, 10% and 30%.
5.4. 71. Tracking_Filter _Bandwidth?
Enables read out of the tracking filter bandwidth.
Example of Use:
Tracking_Filter _Bandwidth?
enables read out of, e.g.:
T_F _8
10%
- For a 10% bandwidth
5.4.72. X_Ramp_Sensitivity
64
Sets the Recorder X Ramp Sensitivity as in Special Function 03.
Example of Use:
or
X_Ramp_Sensitivity
LOgarithmic
- Sets range to logarithmic (0 to 10
V
span)
from 1 to 10000 Hz
X_Ramp_Sensitivity
Linear
- Sets range to linear (0 to 10 V span) from
lower to upper limit