5.4.35. Identify?
Enables read out the instrument identification.
Example of Use:
IDentify?
enables read out of:
B&K
1053
5.4.36. lnput_Sens
Dummy Message, 1053 takes no action when read in. Exists so that it can be read back
to the 1053 without causing an error when accompanying other Messages, e.g. when
reading back the settings, previously read out and stored using the Setup? Message
(section 5.4.61 ).
5.4.37. lnput_Sens?
Enables read out of the input sensitivity, set using the rear panel Input Sensitivity switch
and finely adjusted using the front panel Vibration Meter "Input Sens." function.
Example of Use:
lnput_Sens?
enables read out of, e.g.:
I_S
3 . 96pC/N
- If the rear panel Input Sensitivity switch is set to
"4pCIN", and adjusted to read 3,96pC/N using the
front panel "Input Sens." function.
5.4.38. Max_Levei_No
Selects the number of vibration levels as in the front panel Vibration Meter "Set Levels"
function.
Example of Use:
Max_Levei_No
3
- Sets 3 vibration levels
Notes:
1 to 9 vibration levels can be set if the rear panel Input Sensitivity switch is set to the
"Unit I ms- 2 " settings, otherwise 1 force level is available if set to "Unit I N".
5.4.39. Max_Levei_No?
52
Enables read out of the number of vibration levels.
Example of Use:
Max_Levei_No?
enables read out of, e.g.:
M_L_N
7
- For 7 vibration levels