Cross_ Over _Frequency_< M); Error - BRUEL & KJAER 1053 Instruction Manual

Vibration exciter control
Table of Contents

Advertisement

Notes:
As when activating the front panel function, the result will appear on the 1053 display.
The value can be read out using the Cross_ Over _Frequency_
(
M)? interface Message,
section 5.4.16.
5.4.15. Cross_Over_Frequency_
(
M
)
Sets the selected cross over frequency as in the front panel function Sweep "Cross-
overs" (numerical entry).
Example of Use:
Cross_Over_Frequency_2
780
or
Cross_ Over _Frequency_2
0 . 78kHz
- Sets cross over frequency 2 to 780Hz
Notes:
There are a maximum of 8 cross over frequencies but the number active is N -1 where N
is the maximum number of Vibration Meter Set Levels (see section 5.4.38). It follows that
to turn off the cross over frequencies, N is set to 1.
See Notes in section 5.4.16.
5.4.16. Cross_ Over _Frequency_
(
M
)
?
5.4.17. Error?
46
Enables read out of the frequency value of the selected cross over frequency.
Example of Use:
Cross_Over_Frequency_2?
enables read out of, e.g.:
C_O_F _2
779. 999Hz
Notes:
- If set using the example in section 5.4.15
The resolution of all frequencies is 1,19 mHz. A consequence of this is that about each
sixth value will be rounded down by 1 mHz.
Enables read out of the error number of the last encountered error.
Example of Use:
Error?
enables read out of, e.g.:
ERROR
085
- Assuming the last encountered error was no. 85