Connection Of Heated Line For Portable Version (Part No. 01G6190); Preparations For Commissioning; Commissioning; Figure 10 Heated Line Connection For Portable Version Up To 10 A - M&C CSS-VC1 Instruction Manual

19", wall mount or portable in carrying case with software version
Hide thumbs Also See for CSS-VC1:
Table of Contents

Advertisement

12.4.3 CONNECTION OF HEATED LINE FOR PORTABLE VERSION (PART NO. 01G6190)

If, for the portable version in a carrying case, the option temperature controller for heated line (Part No. 01G6190) is
selected, on the rear of the housing of the CSS-VC.. there is a 7-pin connector for connection of heated lines type
PSP and PSP 4M and PSP 4M-W. These have at the point of connection and termination a 7-pin connector and a
7-pin socket for connection to the portable gas conditioning unit CSS-VC.. and the portable gas sampling probe
type PSP4000H. This is the power supply to the heated line and gas sampling probe using the CSS-VC...
3 Mains
probe
2 Mains N
1 Mains L
Figure 10
Heated line connection for portable version up to 10 A
13

PREPARATIONS FOR COMMISSIONING

Before initial startup, all plant- and process-specific safety measures must be observed. It is mandatory for the
operator to complete the enclosed risk assessment of the product.
The gas exposure risk must be assessed by the operator with regard to the hazards posed by process and calibration
gas and the setup at the installation site (e.g. tubing, system cabinet/container/plant). If the risk assessment reveals
increased exposure hazards, further measures are required.
A visible label must be attached to the installation site in accordance with the risk assessment provided by the
operator.
14

COMMISSIONING

Before commissioning, the plant and process-specific safety measures must to be observed.
Before switching on the power supply, it should be checked again that the operating voltage (see nameplate) and
the power supply voltage correspond!
Warning
www.mc-techgroup.com
4 Mains
probe
5 PT100
6 PT100
Before commissioning exclude that for one channel as well internal as external
a Pt100 is connected.
CSS-VC-1.6 | 1.02.01
23

Advertisement

Table of Contents
loading

This manual is also suitable for:

Css-vc2

Table of Contents