12.2

RELOCATE THE CONNECTIONS TO THE BACK OF THE HOUSING

If, for example with the 19" mounting, mounting conditions require the connections to be moved up in the
connection panel onto the back of the device, this is quite possible:
1. Remove the connection-panel mounting screws (see Figure 3)
2. Remove the cover-bracket mounting screws on the rear panel
3. Mount the connection plate with 2 mounting screws on the rear panel
4. Mount the cover bracket with 2 connection screws on the device cover
12.3

CONNECTION OF THE HEATED LINE WITH ANTIKINK ADAPTER FOR PORTABLE VERSION

(ART. NO. 01G9060 OR 01G9061)
Sample gas IN
bulkhead fitting
Pressure ring 6mm Øi
Support sleeve 4mm
Figure 5
Connecting heated line DN4/6 with antikink adapter
1. Place special adapters according to the above drawing on Teflon tubing;
2. Slide the support sleeve slide into the Teflon tubing;
3. Draw the Teflon tubing all the way into the bulkhead union 'Sample Gas ON' and hand-tighten the
adapter;
4. Tighten the adapter with wrench (SW 14) 1 1/4 turns; here hold the lock nut of the bulkhead union with a
wrench (SW 15);
5. Insert the 10 mm pipe of the heating line all the way into the adapter and hand-tighten using the nut;
6. Tighten nut with wrench (SW 19), 1 1/4 turns; here hold the adapter with wrench;
The union is now cut to be gas-tight and can be loosened as often as required.
www.mc-techgroup.com
Adapter
Clip ring 6mm Øi
Clip ring 10mm Øi
CSS-VC-1.6 | 1.02.01
Junction nut 10mm Øi
Pressure ring 10mm Øi
Heated sample line